Chloride pseudotitration in wines by FIA with a Ag2S/Ag tubular electrode as detector

José L. F. C. Lima*, António O. S. S. Rangel

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

11 Citations (Scopus)
5 Downloads

Abstract

The determination of chloride in wines by potentiometric titration with silver cation in a nonsegmented flow system (FIA) using a tubular electrode of silver/silver sulfide detector is described. There was good agreement between the results obtained using the FIA titration and those obtained using the reference procedure. The methodology developed is adequate for the determination of chloride in various types of wines with chloride contents between 5 and 600 mg/liter, at sampling rates of 60 and 120 samples per hour.
Original languageEnglish
Pages (from-to)356-363
Number of pages8
JournalJournal of Food Composition and Analysis
Volume2
Issue number4
DOIs
Publication statusPublished - Dec 1989

Fingerprint

Dive into the research topics of 'Chloride pseudotitration in wines by FIA with a Ag2S/Ag tubular electrode as detector'. Together they form a unique fingerprint.

Cite this