Flow-through solid-phase based optical sensor for the multisyringe flow injection trace determination of orthophosphate in waters with chemiluminescence detection

Inês P.A. Morais, Manuel Miró, Matias Manera, José Manuel Estela, Víctor Cerdà, M. Renata S. Souto, António O.S.S. Rangel*

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

46 Citations (Scopus)
3 Downloads

Abstract

In this work, a novel flow-through solid-phase based chemiluminescence (CL) optical sensor is described for the trace determination of orthophosphate in waters exploiting the multisyringe flow injection analysis (MSFIA) concept with multicommutation. The proposed time-based injection flow system relies upon the in-line derivatisation of the analyte with ammonium molybdate in the presence of vanadate, and the transient immobilisation of the resulting heteropolyacid in a N-vinylpyrrolidone/divinylbenzene copolymer packed spiral shape flow-through cell located in front of the window of a photomultiplier tube. The simultaneous injection of well-defined slugs of luminol in alkaline medium and methanol solution towards the packed reactor is afterwards performed by proper switching of the solenoid valves. Then, the light emission from the luminol oxidation by the oxidant species retained onto the sorbent material is readily detected. At the same time, the generated molybdenum-blue compound is eluted by the minute amount of injected methanol, rendering the system prepared for a new measuring cycle. Therefore, the devised sensor enables the integration of the solid-phase CL reaction with elution and detection of the emitted light without the typical drawbacks of the molybdenum-blue based spectrophotometric procedures regarding the excess of molybdate anion, which causes high background signals due to its self-reduction. The noteworthy features of the developed CL-MSFIA system are the feasibility to accommodate reactions with different pH requirements and the ability to determine trace levels of orthophosphate in high silicate content samples (Si/P ratios up to 500). Under the optimised conditions, a dynamic linear range from 5 to 50μgPl -1 for a 1.8ml sample, repeatability better than 3.0% and a quantification limit of 4μgPl-1 were attained. The flowing stream system handles 11 analysis h-1 and has been successfully applied to the determination of trace levels of orthophosphate in environmental samples such as mineral, ground, tap and pond waters as well as samples from a water-steam cycle of an incineration plant. The t-test comparison of the means for the developed optical sensor and the molybdenum-blue spectrophotometric APHA/AWWA/WPCF reference method revealed that there is no evidence of significant differences between the obtained results at the 95% confidence level.

Original languageEnglish
Pages (from-to)17-24
Number of pages8
JournalAnalytica Chimica Acta
Volume506
Issue number1
DOIs
Publication statusPublished - 17 Mar 2004

Keywords

  • Chemiluminescence
  • Multisyringe flow injection analysis
  • Orthophosphate determination
  • Solid-phase optical sensor
  • Waters