A flow injection system for the potentiometric determination of chloride (Cl) in vegetables is reported. A tubular chloride ion-selective electrode based on a homogeneous crystalline membrane without inner reference solution was used as indicator electrode. A methodology to minimize the influence of the sample matrix on the electrodes response is described. The results obtained with the developed flow injection manifold are in good agreement with those provided by the reference method and present good precision (relative standard deviations lower than 2%). Sample throughput ranges from 150 to 300 determinations per hour.