T1 - Temperature changes on the root surfaces of mandibular incisors after an 810-nm high-intensity intracanal diode laser irradiation
AU - Alvarez, Andrea da Fonseca
AU - Moura-Netto, Cacio
AU - Frugoli, Alexandre Daliberto
AU - Fernando, Casemiro
AU - Aranha, Ana Cecilia Correa
AU - Davidowicz, Harry
N1 - Copyright:
Copyright 2014 Elsevier B.V., All rights reserved.
PY - 2012/1
Y1 - 2012/1
N2 - Temperature changes caused by laser irradiation can promote damage to the surrounding dental tissues. In this study, we evaluated the temperature changes of recently extracted human mandibular incisors during intracanal irradiation with an 810-nm diode laser at different settings. Fifty mandibular incisors were enlarged up to an apical size of ISO No. 40 file. After the final rinse with 17% ethylenediaminetetraacetic acid, 0.2% lauryl sodium sulfate biologic detergent, and sterile water, samples were irradiated with circular movements from apex to crown through five different settings of output power (1.5, 2.0, 2.5, 3.0, and 3.5 W) in continuous mode. The temperature changes were measured on both sides of the apical and middle root thirds using two thermopar devices. A temperature increase of 7 °C was considered acceptable as a safe threshold when applying the diode laser. Results: The results showed that only 3.5-W output power increased the outer surface temperature above the critical value. Conclusion: The recommended output power can be stipulated as equal to or less than 3 W to avoid overheating during diode laser irradiation on thin dentin walls.
AB - Temperature changes caused by laser irradiation can promote damage to the surrounding dental tissues. In this study, we evaluated the temperature changes of recently extracted human mandibular incisors during intracanal irradiation with an 810-nm diode laser at different settings. Fifty mandibular incisors were enlarged up to an apical size of ISO No. 40 file. After the final rinse with 17% ethylenediaminetetraacetic acid, 0.2% lauryl sodium sulfate biologic detergent, and sterile water, samples were irradiated with circular movements from apex to crown through five different settings of output power (1.5, 2.0, 2.5, 3.0, and 3.5 W) in continuous mode. The temperature changes were measured on both sides of the apical and middle root thirds using two thermopar devices. A temperature increase of 7 °C was considered acceptable as a safe threshold when applying the diode laser. Results: The results showed that only 3.5-W output power increased the outer surface temperature above the critical value. Conclusion: The recommended output power can be stipulated as equal to or less than 3 W to avoid overheating during diode laser irradiation on thin dentin walls.
KW - Diode laser
KW - Endodontics
KW - Intracanal irradiation
KW - Temperature increase
UR - http://www.scopus.com/inward/record.url?scp=84861441831&partnerID=8YFLogxK
U2 - 10.1117/1.JBO.17.1.015006
DO - 10.1117/1.JBO.17.1.015006
M3 - Article
C2 - 22352650
AN - SCOPUS:84861441831
SN - 1083-3668
VL - 17
JO - Journal of Biomedical Optics
JF - Journal of Biomedical Optics
IS - 1
M1 - 015006
ER -