Resumo
A picrate ion-selective electrode without inner reference solution and based on bis(triphenylphosphoranylidene)ammonium picrate dissolved in 2-nitrophenyloctyl ether was constructed and its characteristics assessed. With respect to reducing sugar determination in Port wine, a tubular electrode based on the same sensor system was developed and incorporated into a flow-injection manifold devised for this determination. The methodology involves a reaction between the reducing sugar and picric acid, the decrease in the picrate concentration being monitored with the tubular electrode. A linear dependence between peak width, at a fixed potential and the logarithm of sugar concentration in the range 25-200 g l-1 was obtained. The proposed system allows wine samples to be injected with no previous treatment and allows a sample throughput of 50 determinations per hour, a relative standard deviation less than 4%, and the results are comparable to the reference procedures.
| Idioma original | English |
|---|---|
| Páginas (de-até) | 122-128 |
| Número de páginas | 7 |
| Revista | Analytica Chimica Acta |
| Volume | 308 |
| Número de emissão | 1-3 |
| DOIs | |
| Estado da publicação | Publicado - 9 jun. 1995 |