Dimensions of arachnoid bulk ratio: a superior optic nerve sheath index for intracranial pressure

David Berhanu*, Inês Carneiro, Ana Patrícia Antunes, Luís Abegão Pinto, Isabel Fragata, Joana Tavares Ferreira, Lia Lucas Neto

*Autor correspondente para este trabalho

Resultado de pesquisarevisão de pares

1 Citação (Scopus)

Resumo

Background: Discrepancies in the literature regarding optimal optic nerve sheath diameter (ONSD) cutoffs for intracranial pressure (ICP) necessitate alternative neuroimaging parameters to improve clinical management. Purpose: To evaluate the diagnostic accuracy of the dimensions of the perineural subarachnoid space to the optic nerve sheath ratio, measured using US, in predicting increased ICP. Materials and Methods: In a prospective cohort study from April 2022 to December 2023, patients with suspected increased ICP underwent optic nerve US to determine the dimensions of arachnoid bulk (DAB) ratio and ONSD before invasive ICP measurement. Correlation between the parameters and ICP, as well as diagnostic accuracy, was assessed using area under the receiver operating characteristic curve (AUC) analysis. Results: A total of 30 participants were included (mean age, 39 years ± 14 [SD]; 24 female). The DAB ratio and ONSD were significantly larger in participants with increased ICP (38% [0.16 of 0.42] and 14% [0.82 of 6.04 mm], respectively; P < .001). The DAB ratio showed a stronger correlation with ICP than ONSD (r s = 0.87 [P < .001] vs r s = 0.61 [P < .001]). The DAB ratio and ONSD optimal cutoffs for increased ICP were 0.5 and 6.5 mm, respectively, and the ratio had higher sensitivity (100% vs 92%) and specificity (94% vs 83%) compared with ONSD. Moreover, the DAB ratio better predicted increased ICP than ONSD, with a higher AUC (0.98 [95% CI: 0.95, 1.00] vs 0.86 [95% CI: 0.71, 0.95], P = .047). Conclusion: An imaging ratio was proposed to predict ICP based on the relative anatomy of the cerebrospinal fluid space, demonstrating more accurate diagnosis of increased ICP and a strong correlation with ICP values, suggesting its potential utility as a neuroimaging marker in clinical settings.

Idioma originalEnglish
Número do artigoe240114
Número de páginas7
RevistaRadiology
Volume312
Número de emissão1
DOIs
Estado da publicaçãoPublicado - 9 jul. 2024

Impressão digital

Mergulhe nos tópicos de investigação de “Dimensions of arachnoid bulk ratio: a superior optic nerve sheath index for intracranial pressure“. Em conjunto formam uma impressão digital única.

Citação