Resumo
A sequential injection system for phosphate determination within a wide concentration range was developed for water samples with high salinity levels. The determination is based on molybdenum blue chemistry using the standard addition quantification method. The detection system included a multi-reflective flow cell coupled to a light emitting diode, enabling the minimization of the schlieren effect. The developed system gave a LOD of 0.3 μmol P L -1 and LOQ of 1.1 μmol P L -1 with a sample consumption of 125 μL. A determination rate of about 30 h -1 was obtained, and the developed method was effectively applied to interstitial, transitional and coastal waters.
| Idioma original | English |
|---|---|
| Páginas (de-até) | 1452-1457 |
| Número de páginas | 6 |
| Revista | Analytical Methods |
| Volume | 4 |
| Número de emissão | 5 |
| DOIs | |
| Estado da publicação | Publicado - mai. 2012 |