Simultaneous determination of total iron and chromium(VI) in wastewater using a flow injection system based on the sandwich technique

Alberto N. Araujo*, Jose Luis F.C. Lima, Antonio O.S.S. Rangel, Julian Alonso, Jordi Bartroli, Raquel Barber

*Autor correspondente para este trabalho

Resultado de pesquisarevisão de pares

29 Citações (Scopus)

Resumo

A flow injection system for the simultaneous determination of total Fe and CrVI based on the sandwich technique is described. Both metals were determined spectrophotometrically at 526 nm with a single detector over the range 0.2-10 p.p.m. The sampling frequency achieved was 60 samples h -1 and the standard deviations were 0.09 and 0.08 p.p.m. for total Fe and CrVI, respectively.
Idioma originalEnglish
Páginas (de-até)1465-1468
Número de páginas4
RevistaThe Analyst
Volume114
Número de emissão11
DOIs
Estado da publicaçãoPublicado - 1989
Publicado externamenteSim

Impressão digital

Mergulhe nos tópicos de investigação de “Simultaneous determination of total iron and chromium(VI) in wastewater using a flow injection system based on the sandwich technique“. Em conjunto formam uma impressão digital única.

Citação