TY - JOUR
T1 - Turbidimetric determination of chloride in different types of water using a single sequential injection analysis system
AU - Mesquita, Raquel B. R.
AU - Fernandes, Sílvia M. V.
AU - Rangel, António O. S. S.
PY - 2002/7/2
Y1 - 2002/7/2
N2 - A sequential injection analysis system for the turbidimetric determination of chloride in different types of water is proposed. The determination is based on the reaction of chloride with silver ions and the subsequent measurement of the turbidity caused by silver chloride precipitation. In this method, the use of toxic reagents, such as mercury thiocyanate, commonly employed in most spectrophotometric techniques for chloride determination, is avoided. The main feature of the developed system is the use of a single configuration to carry out the determination over a wide concentration range (2-400 mg L-1) by changing only the aspirated sample volume. This characteristic allows the determination of chloride in ground, surface and wastewaters using the same manifold. In addition, a considerable saving of precipitating reagent is achieved due to non-continuous consumption. The results obtained with the developed system were statistically indistinguishable from those of the potentiometric titration reference method. Relative standard deviations for ten consecutive injections were lower than 3.7%, with a sampling frequency of between 55 and 57 determinations per hour.
AB - A sequential injection analysis system for the turbidimetric determination of chloride in different types of water is proposed. The determination is based on the reaction of chloride with silver ions and the subsequent measurement of the turbidity caused by silver chloride precipitation. In this method, the use of toxic reagents, such as mercury thiocyanate, commonly employed in most spectrophotometric techniques for chloride determination, is avoided. The main feature of the developed system is the use of a single configuration to carry out the determination over a wide concentration range (2-400 mg L-1) by changing only the aspirated sample volume. This characteristic allows the determination of chloride in ground, surface and wastewaters using the same manifold. In addition, a considerable saving of precipitating reagent is achieved due to non-continuous consumption. The results obtained with the developed system were statistically indistinguishable from those of the potentiometric titration reference method. Relative standard deviations for ten consecutive injections were lower than 3.7%, with a sampling frequency of between 55 and 57 determinations per hour.
UR - http://www.scopus.com/inward/record.url?scp=0036088451&partnerID=8YFLogxK
U2 - 10.1039/b200456a
DO - 10.1039/b200456a
M3 - Article
C2 - 12094945
AN - SCOPUS:0036088451
SN - 1464-0325
VL - 4
SP - 458
EP - 461
JO - Journal of Environmental Monitoring
JF - Journal of Environmental Monitoring
IS - 3
ER -